Here are the changes related to the presets used by the SmartNormalize algorithm:


Initial Scanner

...

Settings - Files Update

Warning
titleAttention

For proper functioning of the system, it is required that you download and use the new versions of files.

Initial Scanner Settings - Files Update

The files with the initial scanner settings have been updated. You can download the updated files from the page: Initial Scanner Settings

...

  • For the HP_1 boards:
    • The Period has been changed (HPODrivers_Shadow_HeliumSpeed.ini)
    • Acceleration presets have been added (HPODrivers_Shadow_HeliumSpeed.ini)
    • hws_prosilica.xml has been added - contains settings for HPO Shadow with the Prosilica camera. Now includes two additional elements (by default commented out and thus inactive):
      • Under <benchmark_fps>, <!--measure/--> - uncomment to <measure/> to enable the benchmark measurement of the maximum FPS
      • <!--packetSize>8228</packetSize--> - for some installations, the automatically set packet size causes video artifacts like missing lines - in that case, uncomment to <packetSize>8228</packetSize> and decrease the initial "8228" until artifacts are gone.
  • For the rest .ini files
    • Unused blocks have been removed
    • Some elements have been renamed 

HPPacor - Folder Update

The content of the HPPacor folder:

  • \Dropbox (OctoNus)\Download\HP\HP_For_Calibration_11D_22.5D_HP_SMC_v2_Ximea

...has been updated.

For the initial scanner alignment in HPPacor for Ximea v2, you must use the updated files.

...


Hasp HL Driver - Update

After installation of HP Oxygen of version 5.2.13 or later, on running the system you may obtain the error message:

...